Front Panel Operation
Press CONFIG + ACV button. Use and buttons to locate “BAND
WIDTH” submenu, and then press ENTER to select it. Look for the desired
bandwidth using and buttons. Then press ENTER on your
selection. There are three options: 3Hz, 20Hz and 200Hz. The locations
of these buttons are shown with red rectangle frames in Figure 4-2.
Figure 4-2
Remote Interface Operation
From your PC terminal, use the following command to specify the filter
type:
DETector:BANDwidth {3|20|200|MIN|MAX}
4.1.2.2 Digital Filter
Definition:
1201 uses an averaging digital filter to yield a reading for display from a
specified number of measurement readings in the past. The past
measurement readings are stored in a stack of memory. The number
may be in the range of 2 to 100. User may select one of the two modes
of digital filter operations: Moving Average and Repeating Average.
The moving average filter puts the specified number of reading
conversions in a first-in, first-out order. The very first measurement
reading simply fills up the stack. To yield a reading for display, the filter
produces an average of the stacked measurement readings every time a
new measurement reading is available and replaces the oldest reading in