6 Error Messages
Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. When user has read all errors
from the queue, the ERROR annunciator turns off. 1201 beeps once
each time an error occurs.
Should more than 20 errors have existed, the last error stored in the
queue (the most recent error) is replaced with -350, “Too many errors”.
No additional errors are saved until you remove errors from the queue,
the multimeter responds with +0, “No error”.
The error queue is cleared when power has been off or after a *CLS (clear
status) command has been executed. The *RST (reset) command does
not clear the error queue.
6.1 Error Type
There is only one type over here, “execution errors”. If the errors user
can’t find here please contact with your seller for help.
6.1.1 Execution Errors
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string.
-103 Invalid separator
An invalid separator was found in the command string.