Bnc 1201 - 6 1/2 Digit Digital Multimeter Manuel d'utilisateur Page 147

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 177
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 146
148
[SENSe:]TEMPerature:RTD:ALPHa? [MINimum|MAXimum]
Query the alpha constant for the user type.
[SENSe:]TEMPerature:RTD:BETA {<value>|MINimum|MAXimum}
Set the beta constant for the user type.
[SENSe:]TEMPerature:RTD:BETA? [MINimum|MAXimum]
Query the beta constant for the user type.
[SENSe:]TEMPerature:RTD:DELTa {<value>|MINimum|MAXimum}
Set the delta constant for the user type.
[SENSe:]TEMPerature:RTD:DELTa? [MINimum|MAXimum]
Query the delta constant for the user type.
[SENSe:]TEMPerature:SPRTD:RZERo {<value>|MINimum|MAXimum}
Set the sensor R value at 0 degree Celsius.
[SENSe:]TEMPerature:SPRTD:RZERo? [MINimum|MAXimum]
Query the sensor R value at 0 degree Celsius.
[SENSe:]TEMPerature:SPRTD:A4 {<value>|MINimum|MAXimum}
Set the A4 coefficient.
[SENSe:]TEMPerature:SPRTD:A4? [MINimum|MAXimum]
Query the A4 coefficient.
[SENSe:]TEMPerature:SPRTD:B4 {<value>|MINimum|MAXimum}
Set the B4 coefficient.
[SENSe:]TEMPerature:SPRTD:B4? [MINimum|MAXimum]
Query the B4 coefficient.
[SENSe:]TEMPerature:SPRTD:AX {<value>|MINimum|MAXimum}
Set the A coefficient.
[SENSe:]TEMPerature:SPRTD:AX? [MINimum|MAXimum]
Vue de la page 146