Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS Page -30 / 36 -
5.2. Example 2
Realize in local mode a frequency measurement on start input
Same initial set-up and test signal as for example n°1
Setting up the instrument
Use the front panel keyboard to reset the Setup menu as follows
Start trigger External
Start level 1.50 V
Start polarity +
Start the measurement
Display the result line frequency, and press the OK keys to start the measurement
The front panel leds busy must be on and tilt each second
Result must be
Frequency = 100 Hz
For stopping the measurement press the OK key