Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS Page -24 / 36 -
STAT
Syntax: STAT CLEAR
STAT? POW|TRIG|CLK|TEMP|SCAN
Form: Set & Query
Description: Clear or query the status.
Command
Parameter:
CLEAR clear the trig status
SCAN, this command measure all analog values, convert analog
value in digital value, save the digital value in the interface
controller and refresh the value in the front panel
Query
Parameter
POW return the power levels (+6V, -6V, +3.3V, +1.8V, +12V)
TRIG return the trigger status
CLK return the clock status
TEMP return the equipment temperature
Example: STAT CLEAR
STAT SCAN
STAT? POW => :STAT POW,6.00,-6.00,3.30,1.80,12.00
STAT? TRIG => :STAT TRIG,1,1
STAT? CLK => :STAT CLK,INTERNAL
STAT? TEMP => :STAT TEMP,45.27