Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS Page -23 / 36 -
RUN
Syntax: RUN DELAY|FREQ|COUNT
RUN?
Form: Set & Query
Description: Start measure or ask if measuring.
Command
Parameter:
DELAY start a delay measurement
FREQ start a frequency measurement
COUNT start a count measurement
Query response :RUN YES measurement running
:RUN CAL calibration running
:RUN NO no measurement running
Example: RUN DELAY
RUN? => :RUN YES
SAMPLES
Syntax: SAMPLES <samples>
SAMPLES?
Form: Set & Query
Description: Set or query the measurement samples.
Command
Parameter:
<samples> number of samples
Specified limit 1 <= samples <= 1000
Example: SAMPLES 500
SAMPLES? => :SAMPLES 500